may-quang-pho-uv-730-3

Spectral machine UV730

Model: UV730

Manufacturer: Jasco

Spectral machine UV730 the airline's Jasco with user-friendly features and safety with the use of the optical system and two beam giving back the correct results. Flexible operation system with 2 the interface: LCD color touch screen and software on your computer.

Order

Why shopping at INNOTEC?

  1. Get every day.
  2. View restaurant > Satisfaction > Payments
  3. Be paid under the policy change preferences
  4. Guarantee 12 month, support 24/24

SKU: UV730 Categories: ,

Featured specification of spectral machine UV730

  • Friendly and safe features with users.
  • Optical system with two beam giving back the correct results
  • The collection of data calculation is fully automated with star IQ, make the user reduce the operation and the process took place exactly ,Trust.

Spectral machine UV730 two integrated operation interface:

  • Through an LCD color touch screen and through the software Spectra Manager II on a computer (( the latest version was developed by JASCO).
  • Both interfaces are precise control allows to control the whole system and process data accurately. Users can easily use this interface with the interface.

may-quang-pho-uv730

  • The size of the compact UV730 spectrum: 441 (In )x468 (D )x 216(H ) mm, space-saving lab.
  • High scan speed up to 8000nm/min without error in the measuring process is by spectroscopy equipped fast response detectors.
  • Offer comes including accessories temperature control and sampling software packages are available to optimize the application of UV730.
  • Model UV730 in accordance with the regulations of GMP/GLP laboratory.
may-quang-pho-uv730-2
The console software via your computer

Learn more about the product line spectroscopy UV-Vis in list: UV spectroscopy- Vis

 

 

There are no reviews yet.

Be the first to review "the spectrum UV730”

The criteria

Value description

Optical system

Type 2 beam

Monochromatic beam

Monochromatic lights layout according to the circle

The light source

Halogen Lamps, Deuterium lamps

The length of the wavelength range

190 – 1100 nm

Accuracy of wavelength

±0.2nm( in 656.1 nm)

Repeat mode of wavelength

±0.1nm

Spectral bandwidth(SBW)

1 nm

Dispersion

1%(198NM solution of KCL 12 g/L )

0.02%(220 NM solution of NaI 10 g/L)

0.02%(340 NM solution of NaNO2 50g/L)

0.02%(370 NM solution of NaNO2 50g/L)

SBW:1 nm

The measuring range of the spectrum

-3 to 3 Abs

The precision of the light

+/-0.0015 Abs (0 to 0.5 Abs)
+/-0.0025 Abs (0.5 to 1 Abs)
+/-0.3 %T Tested
with NIST SRM 930D

Speed of light

+/-0.0005 Abs (0 to 0.5 Abs)
+/-0.0005 Abs (0.5 to 1 Abs)
Tested with NIST SRM 930D

Scan speed

10-8000NM/minute

Rotation speed

24000 NM/minute

RMS noise

0.00004 Abs

(0 Abs, the wavelength: 500 nm, measurement time: 60 sec, SBW: 1 nm)

Stability of the beam source

0.0004 Abs/h

(The value obtained more than an hour after powering,When the room temperature stable, the wavelength : 250 nm ,response : slow)

 

The flat-beam sources

+/-0.0005 Abs (200 – 1000 nm)

The sensor

Silicon photodiode

Standard tools

IQ accessories, Start button, Analog output

Standard program

ABS/%T meters, Quantitative analysis, Spectral measurement,
Cycle time measurement, Fixed wavelength measurement, Validation,
Daily maintenance, Macro Commands (only used for iRM interface),
Calculate the time for measuring two wavelengths (only for PC)

Size and weight

486(In) x 441(D) x 216(H) mm, 15 kg

The source of the request

120VA

Operating requirements

In room temperature:15-30theC, humidity: <85%